Product complaint number: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: one open box with twenty-one unopened samples of product code: k833 was returnedfor evaluation.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no anomalies were observed during evaluation.Functional test was performed, and the pull force result were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, and the following was obtained: how many needles popped off the suture during this single procedure? no specifics given.There are 15 needles missing from the original box.Was there any adverse patient outcome? no harm done to patient.Procedure name and date? robotic colectomy (gastrojejunostomy) tube replacement event date? on (b)(6) 2022.Are any samples available of evaluation? please provide facility contact person¿s name, mailing address and telephone.A shipper kit will be sent to help return the product to us.Yes.
|