Unomedical reference number: (b)(4).Event occurred in the united states.On (b)(6) 2022, it was reported that the patient's infusion set's tubing came apart close to the site location at all different times which led to high blood glucose levels.The site location was left, and right side of patient's abdomen and the pump was also on the same side in relation to the site location.The infusion had been used for first or second day.Reportedly, the infusions were not stored, or used, in a place where they might have been exposed to extreme temperatures and humidity.There was no stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.Currently, her blood glucose level was between 261-315 mg/dl.No further information available.
|