| Model Number |
AVSM10060 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned to the manufacturer for evaluation and images were not provided for review.Based on information available and because the sample was not available for evaluation the reported issue is closed with inconclusive evaluation result.A definite root cause could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use states "prior to covered stent deployment, unlock the red safety lock (figure 7) by pressing down and pulling it back towards the end of the grip from the locked position into the unlocked position.Ensure that the red safety lock is completely retracted and that the symbol for the unlocked position is fully visible." expiry date: 10/2023.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the device allegedly had a activation failure.There was no reported patient injury.
|
| |
|