• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS VENACURE NEVERTOUCH DIRECT PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS VENACURE NEVERTOUCH DIRECT PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number 11403101
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
An end user reported an issue with a pvak -- 400 micron perforator fiber.When the physician attempted to insert the fiber into the machine, it broke in half.A new fiber was opened to complete the procedure.The patient did not experience any adverse effects, harm, or require medical intervention because of this incident.
 
Additional Manufacturer Narrative
The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
Returned for evaluation was a fiber noted to be coiled and in 2 pieces.A visual inspection noted that the fiber fracture it was located at 46.7cm from the distal end of the fiber shaft.The break point in the fiber glass core is not jagged and the break appears to have occurred under a low stress condition.Since the beak appears to have occurred under a low stress condition, the stretched / distorted buffer on one side of the break is most likely due to handling after the break occurred.The customer's reported complaint description of" the physician attempted to insert the fiber into the machine, it broke in half" was confirmed based on evaluation of the returned fiber sample.Although confirmed, a definitive root cause for the fiber fracture could not be determined.Since the break was confined to a localized point along the fiber and there was no apparent buffer damage beyond the break point, general mishandling does not appear to be associated with the break.There is no way to definitively determine what exactly caused the break but it is possible that bending of the fiber when being connected to a laser in combination with some material variability in the fiber glass core may have resulted in the break that occurred.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference pr(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENACURE NEVERTOUCH DIRECT PROCEDURE KIT
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15173499
Report Number1319211-2022-00033
Device Sequence Number6213043
Product Code GEX
UDI-Device IdentifierH787114031015
UDI-PublicH787114031015
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K171921
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number11403101
Device Lot Number5723653
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer08/24/2022
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-