MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM 04.01.0003 STD HUMERAL DIAPHYSIS - CEMENTLESS - 8; SHOULDER HUMERAL DIAPHYSIS - CEMENTLESS Back to Search Results
Model Number 04.01.0003
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 12 july 2022.Lot 176935: (b)(4) items manufactured and released on 13-nov-2017.Expiration date: 2022-oct-16.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Other device involved: batch review performed on 12 july 2022.Reverse shoulder system 04.01.0172 glenosphere 36xø27 (k170452) lot 174733: (b)(4) items manufactured and released on 26-sep-2017.Expiration date: 2022-sep-13.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.
 
Event or Problem Description
Revision surgery performed 4 years after the primary surgery due to glenosphere position too superior (impingement) and stem was not stable.All implants revised.The patient already had previous revision surgery due to infection.
 
Additional Manufacturer Narrative
Clinical evaluation performed by medical affairs department: revision 4 years after the primary rsa surgery in a female patient.The patient already had previous revision surgery due to infection.This may have created a difficult situation that made a superior position of the glenosphere/baseplate inevitable; a possible consequence of this could be impingement and subsequent mobilization of the humeral stem.All implants were revised.Due to the low quality of the images it is not possible to confirm if the position of the glenosphere is too superior, however, the humeral stem appears to have shifted laterally in the canal.We cannot tell if this is the result of post-surgery migration or the surgeon's choice: in the latter case, no explanation for this decision was supplied.In these conditions, it is very likely that due to the impingement, the stem moved from its original position and it results as shown in the radiographs.No reason to suspect faulty implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM 04.01.0003 STD HUMERAL DIAPHYSIS - CEMENTLESS - 8
Common Device Name
SHOULDER HUMERAL DIAPHYSIS - CEMENTLESS
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15174131
Report Number3005180920-2022-00584
Device Sequence Number17714277
Product Code PHX
UDI-Device Identifier07630040706025
UDI-Public07630040706025
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/16/2022
Device Model Number04.01.0003
Device Catalogue Number04.01.0003
Device Lot Number176935
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-