• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SENSATION SHORT THROW; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SENSATION SHORT THROW; SNARE, FLEXIBLE Back to Search Results
Model Number M00562691
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a sensation small oval flexible snare was used during a procedure performed on unknown date.During the procedure and inside the patient, the wire unraveled.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.
 
Additional Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: problem code a0401 captures the reportable event of snare loop break.Block h10: investigation results a sensation small oval flexible snare was received for analysis.Visual and microscope inspection of the returned device revealed that the loop was partially broken.No other device problems were noted.The reported event of loop break was confirmed.During product analysis it was observed that the loop was partially broken.An investigation to address this problem has been completed.The investigation determined that the most assignable cause is human error as it was determined that the product builder did not follow the procedure for inspection of the loop as the defect was not discarded and the unit was sold.The investigation concluded that this is considered an isolated event.Based on the information available and the returned device analysis, the most probable root cause for the reported complaint is manufacturing deficiency.
 
Event or Problem Description
It was reported to boston scientific corporation that a sensation small oval flexible snare was used during a procedure performed on unknown date.During the procedure and inside the patient, the wire unraveled.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION SHORT THROW
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15174243
Report Number3005099803-2022-04271
Device Sequence Number12042482
Product Code FDI
UDI-Device Identifier08714729158158
UDI-Public08714729158158
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K131700
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2025
Device Model NumberM00562691
Device Catalogue Number6269
Device Lot Number0029083544
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date09/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-