| Model Number |
515105 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No photos or physical samples that display the reported condition were available for investigation.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device.As the lot involved in this incident is unknown, a device history review cannot be performed, and additional retained samples cannot be evaluated.Based on the available information we are not able to identify a root cause at this time.The m12 assembly fixture is recommended to ease the connection of the protector to the vial.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd phaseal¿ protector p50 was leaking.The following information was provided by the initial reporter: during preparation in the isolator, the phaseal protector system did not fit on the cyclophosphamide vial.The cyclophosphamide leaked (spray) into the isolator.
|
| |
|
Search Alerts/Recalls
|