EXACTECH, INC. ALT HA S CLR STD SZ 5; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS
|
Back to Search Results |
|
| Model Number |
190-30-05 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Implant Pain (4561); Aseptic loosening (4883)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, this female patient had a right side tha on (b)(6) 2017.The patient was brought back for right hip pain.The hip was dislocated, the femoral head removed.The stem was checked, found to be loose and removed.Attention turned to cup.The liner was removed with the use of a bone screw.The cup was checked, found to be well fixed.A new g1 32 lip xle liner was implanted.Attention turned back to the femur where a new size 6 std offset hps stem was cemented in place along with a new 32 +3.5 biolox head.There is no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.The devices are not returning as the patient requested them.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant medical products: 130-32-51 nv gxl liner neutral, 32mm id group 1 cups - 4120690; 170-32-03 biolox delta femoral head 32mm od, +3.5mm- 4320796.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of an insufficient bond between the femoral stem and the bone, an under sized femoral stem for the patient¿s anatomy/femoral bone, patient related conditions, or a combination of these possibilities, which led to aseptic (non-infected) femoral stem loosening and pain.However, this cannot be confirmed as the devices were not available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d8, f10 (remove information erroneously added on initial report), h6 - health effect - clinical code, component code, type of investigation, investigation findings, investigation conclusion and h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.A definitive root cause was unable to be determined; however, may have resulted from an insufficient bond between the femoral stem and the bone, an under sized femoral stem for the patient¿s anatomy/femoral bone, patient related conditions, or a combination of these possibilities, which led to aseptic (non-infected) femoral stem loosening and pain.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|