• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD PRIMARY GRAVITY IV ADMINISTRATION SETS; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD PRIMARY GRAVITY IV ADMINISTRATION SETS; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number MX9128
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: a complaint of a misassembled set was received from the customer.A sample was received for investigation.Through visual inspection the customer complaint was confirmed.The roller clamp was assembled upside down.A device history record review for model mx9128 lot number 22059056 was performed.The search showed that a total of (b)(4) units in 2 lot numbers were built on 05may2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The manufacturing plant was notified of this defect.The root cause was determined to be an error during assembly.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that the bd primary gravity iv administration sets experienced tubing defective.The following information was provided by the initial reporter: we had an mx9128 iv set - lot number 22059056 that when we tried to use it the roller clamp was upside down on the tubing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRIMARY GRAVITY IV ADMINISTRATION SETS
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15174885
Report Number9616066-2022-01061
Device Sequence Number17709620
Product Code FPA
UDI-Device Identifier50885403237629
UDI-Public50885403237629
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMX9128
Device Catalogue NumberMX9128
Device Lot Number22059056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-