MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PEEK 12 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE
|
Back to Search Results |
|
Model Number 228151 |
Device Problems
Device-Device Incompatibility (2919); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported by a healthcare professional in chile that during a knee arthroscopy with meniscal repair procedure on (b)(6) 2022, it was observed that the peek implant did not come out of the gun when the trigger was pressed on the truespan meniscal repair system peek 12 degree device.According to the report, the implant was found to be trapped in the sheath when the white sheath was opened.Another like device was used to complete the procedure with a delay of 15 minutes.There were no adverse patient consequences reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch wll be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation, however a video and 2 photos were provided.Upon reviewing the video and the photos, the second plate is stuck inside the plate container.A manufacturing record evaluation was performed for the finished device 9l22275 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the result of the video and photos review, this complaint can be confirmed.Multiple factors are associated with this type of failure, the complaint device needs to be received in our evaluation lab.Hands on analysis will provide enough evidence to determine why the customer experienced the failure.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available or the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the complaint device was received and evaluated; on visual inspection, the plastic sleeve was removed already removed by the customer.The second plate was found inside the applier needle and is stuck with the tip of the spring pusher.Rests of biological matter can be observed right at the tip of the spring pusher.After unlocking the plate, the trigger was tested several times, it performed as intended, the spring pusher has no structural anomalies.A manufacturing record evaluation was performed for the finished device 9l22275 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection and the functional test result, this complaint can be confirmed.A possible root cause can be attributed to a portion of tissue that was entered inside the applier needle causing a mechanical obstruction of the first plate when it was going to be deployed; this obstruction and the force applied to the trigger caused the spring pusher to slip below the plate and created a jamming condition between the second plate and the spring pusher, however this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|