U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD BUZZ NONSTICK: LENGTH: 21.6CM (8.50), TIP SIZE:; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD BUZZ NONSTICK: LENGTH: 21.6CM (8.50), TIP SIZE:; N/A Back to Search Results
Model Number NHI15
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
This is 1 of 2 reports for the bipolar forceps reportedly used during this event and is linked to mfg report number 3014334038-2022-00171.A facility reported that during an ortho/spine procedure, it was noted that the bipolar forceps/buzz nonstick (nhi15) were no longer conducting at the tips.It was reported that there was a delay of 30 minutes to an hour because the staff had to open replacement instruments.The procedure was completed with replacement instruments.No patient injury occurred.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 the buzz nonstick (nhi15) was returned for evaluation: device history record (dhr): the dhr was reviewed and no abnormalities related to the reported issue were identified.The reported complaint was confirmed.Visual evaluation noted that the returned buzz nonstick forceps were in used condition with the tips tarnished and the insulation discolored due to wear/sterilization issues.Damaged tips and insulation may have interfered with conductivity.No manufacturing, workmanship, or material deficiency has been identified.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUZZ NONSTICK: LENGTH: 21.6CM (8.50), TIP SIZE:
Common Device Name
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15175115
Report Number3014334038-2022-00170
Device Sequence Number12020653
Product Code GEI
UDI-Device Identifier10381780149194
UDI-Public10381780149194
Combination Product (Y/N)N
PMA/510(K) Number
K051429
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNHI15
Device Catalogue NumberNHI15
Device Lot NumberAN2012
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-