| Model Number |
G5S-01A |
| Medical Device Problem Codes |
High impedance (1291); Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device did not detect the attached electrode pads and failed for high impedance.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device activity log.However, the device was put through extensive testing including functionality testing, bench handling, and fast testing without duplicating the report.Internal inspection of the device found no discrepancies.The electrode connector was replaced as a precaution.The device was recertified and returned to the customer.The electrode pads used at the time of the report were not returned as part of this investigation.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|