| Model Number |
0998-00-3023-83 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit is not functional.It is unknown under which circumstances the event occurred or if there was any patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Getinge field service engineer (fse) confirmed reported issue.Display would not work.Replaced the video monitor panel (0160-00-0113) and video cable (0012-00-1747).Verified unit calibrated and passes all functional and safety tests per factory specifications, returned to customer and cleared for clinical use.
|
| |
|
Event or Problem Description
|
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit is not functional.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|