• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR(S); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR(S); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MZ9277
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for additional lot number is as follows: medical device lot #: 20036653.Medical device expiration date: 20-mar-2023.Device manufacture date: 19-mar-2022.Medical device lot #: 22025498.Medical device expiration date: 10-feb-2025.Device manufacture date: 08-feb-2022.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that 6 bd maxzero multi-fuse extension set with needleless connector(s) experienced leakage.The following information was provided by the initial reporter: we had two maxzero iv connectors fail in june.There was a leak at the max zero hub, and needed to be changed three times.
 
Additional Manufacturer Narrative
H6: investigation summary: samples were received and investigated by our quality team.No leak was realized with multiple attempts infusing saline solution with both syringe injection and infusing through other devices and no issue was detected.Due to no leakage, the complaint cannot be verified and the root cause remains unknown.The failure was due to the connected device when the leakage occurred.A device history record review was performed on material mz9277 for lots 20036653 (30mar2020 (b)(4) units produced) and 22025498 (10feb2022 (b)(4) units produced).There were no issues or quality notifications during the production of these sets.H3 other text : see h10.
 
Event or Problem Description
It was reported that 6 bd maxzero multi-fuse extension set with needleless connector(s) experienced leakage.The following information was provided by the initial reporter: we had two maxzero iv connectors fail in june.There was a leak at the max zero hub, and needed to be changed three times.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR(S)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15175226
Report Number9616066-2022-01065
Device Sequence Number6288771
Product Code FPA
UDI-Device Identifier10885403240799
UDI-Public10885403240799
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMZ9277
Device Lot NumberSEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-