Model Number SN60WF
Medical Device Problem Code Material Opacification (1426)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138); Visual Disturbances (2140)
Date of Event 06/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that following an intraocular lens (iol) implant procedure, 5-7 years ago developed blurry vision, and was noted to have a myopic shift in her left eye that has now stabilized.She had worsening glare as well.Her vision decreased to about 20/30 even with the new refraction.On exam it looks like she has a thickened and somewhat opacified optic.There was no trauma or other unusual finding on exam.Additional information was requested and received stating the lens was explanted.
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens was returned cut into four pieces, typical of insertion and removal from the eye.Cracks are observed on one of the pieces of lens.The lens was cleaned for further evaluation.The lens appears clear.A power and resolution inspection could not be conducted due to extensive damage.Product history records were reviewed and the documentation indicated the product met release criteria.The root cause for the reported events could not be determined.A malfunction has not been indicated or information provided that the lens was the cause of the event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15175361
Report Number1119421-2022-01718
Device Sequence Number14710978
Product Code HQL
UDI-Device Identifier00380655093153
UDI-Public00380655093153
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.180
Device Lot Number12385004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Initial Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2015
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexFemale
-
-