|
|
| Model Number |
TSVTH10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implants on teeth sites #19 and #20 were removed due to peri-implantitis.Symptoms as a result of the event: bone loss, infection, inflammation.
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event.Please see associated report: 0002023141-2022-01970.Zimmerbiomet complaint number (b)(4).Patient identifier unknown / not provided.Age and date of birth unknown / not provided.Weight unknown / not provided.Date of event unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformance's affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Search Alerts/Recalls
|
|
|