| Model Number |
K872H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown dental surgery on (b)(6) 2022 and the suture was used.During surgery, the suture was detached from the needle easily during interrupted suturing on oral.It was not a control release needle.There were no adverse patient consequences reported.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 08/24/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 evaluation: the product was returned for evaluation.Visual inspection evaluation was conducted on the returned device.One empty labeled winding former, one detached needle, and one dispensed suture of product were returned for evaluation.Upon visual inspection of the one opened sample, it was observed the needle was broken at the swage area and a part of the needle still was attached to the suture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/16/2022.H3 investigational summary: the product received for analysis was identified as product code k872h.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|