• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL¿ CONNECTOR LUER LOCK (C35); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL¿ CONNECTOR LUER LOCK (C35); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 515200
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that at least one bd phaseal¿ connectors luer lock (c35) leaked between the injector and connector.The following information was provided by the initial reporter: the nurse detached the chemo, turned around and disposed the bag and the patient caught our attention a minute later because blood was running back out of the female phaseal attachment from her line.It seems to only be happening with long-term hookups, so our 24 hour chemotherapy¿s.When nurses change out the bags the female port starts to leak after disconnection sometimes, almost like when the male end is removed the hub on the line remains engaged instead of closing.Again, this has only been observed with 24 hour chemo¿s.
 
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6: investigation summary no photos or physical samples that display the reported condition were available for investigation.A device history review was performed for lot 2010216, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device, no issues related to the reported incident were found for the repotted lot.Five retained samples of the same lot were used for additional evaluation.All product was visually inspected and no defects were observed.Product could be properly connected to mating components and product was verified to meet required specification.Based on the available information we are not able to identify a root cause related to the manufacturing process at this time.H3 other text : see h10.
 
Event or Problem Description
It was reported that at least one bd phaseal¿ connectors luer lock (c35) leaked between the injector and connector.The following information was provided by the initial reporter: the nurse detached the chemo, turned around and disposed the bag and the patient caught our attention a minute later because blood was running back out of the female phaseal attachment from her line.It seems to only be happening with long-term hookups, so our 24 hour chemotherapy¿s.When nurses change out the bags the female port starts to leak after disconnection sometimes, almost like when the male end is removed the hub on the line remains engaged instead of closing.Again, this has only been observed with 24 hour chemo¿s.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL¿ CONNECTOR LUER LOCK (C35)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP  22520
Manufacturer (Section G)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP   22520
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15177046
Report Number3002682307-2022-00204
Device Sequence Number6330024
Product Code ONB
UDI-Device Identifier30382905152000
UDI-Public30382905152000
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K181221
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number515200
Device Catalogue Number515200
Device Lot Number2010216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer10/13/2022
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-