| Model Number |
DL-330-3.5-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) the explanted device was not returned to neuropace for investigation.
|
| |
|
Event or Problem Description
|
|
The treating center identified unusual impedances on the lead during a patient ecog review.When the patient's head was palpated; there was visible artifact observed on the ecog.During the neuropace investigation of the ecog data, we observed artifacts and high impedances on lead 2, indicative of a lead break.The lead was explanted and replaced on (b)(6) 2022.The lead was secured with a dog bone and lead protector.Lead was implanted for 0.6 years.There was no identified cause of the lead break.
|
| |
|