| Catalog Number |
07.02100.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3012447612-2022-00186.
|
| |
|
Event or Problem Description
|
|
It was reported that the pivot screw closest to the tip of two vitality rod grippers broke intra-operatively, making them unusable.The polaris grippers were used to complete the surgery without patient impacts.This is report one of two for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that the pivot screw closest to the tip of two vitality rod grippers broke intra-operatively, making them unusable.The polaris grippers were used to complete the surgery without patient impacts.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned devices match the information in the complaint file and were examined.Visual inspection revealed that the middle screw was missing from both devices.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Device usage: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
|
| |
|