• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3662
Device Problem Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388)
Event Date 07/14/2022
Event Type  Injury  
Event Description
It was reported the patient had shoulder surgery where the device was not placed in surgery mode.Subsequently, the patient has been unable to connect to the ipg and the patient controller is displaying an error message since the procedure.Troubleshooting was attempted by the company representative to no avail.Surgical intervention took place wherein the ipg was explanted and replaced to address the issue.
 
Manufacturer Narrative
Date of event is estimated.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Manufacturer Narrative
The reported observation of ¿no ipg communication¿ was confirmed.The analysis results concluded the inability to establish communication between the programmer and the implantable pulse generator (ipg) was due to the device entering the service application state.The root cause of the device entering the service application state was due to the use of an electro-surgical device during the surgical procedure.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.Per the clinicians manual: per clinicians manual (b)(4) - under warnings - there is sufficient documentation concerning the use of use short-wave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system, and medical procedures (such as therapeutic radiation and electromagnetic lithotripsy).Per the clinicians manual electrosurgery.To avoid harming the patient or damaging the neurostimulation system, do not use monopolar electrosurgery devices on patients with implanted neurostimulation systems.Before using an electrosurgery device, place the device in surgery mode using the patient controller app or clinician programmer app.Confirm the neurostimulation system is functioning correctly after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15179778
MDR Text Key297404343
Report Number1627487-2022-04294
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020222
UDI-Public05415067020222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/24/2019
Device Model Number3662
Device Catalogue Number3662
Device Lot Number6064474
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Treatment
SCS ANCHORS X 2; SCS LEADS X 2
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight68 KG
-
-