• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ AP ID BROTH; CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ AP ID BROTH; CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Catalog Number 448012
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd phoenix¿ ap id broth biological contamination was observed.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer that they started seeing purity plates with pseudomonas aeruginosa.Customer did decontamination and cultured the bag.The bag grew p.Aeruginosa.Customer did additional testing and found the lot to be contaminated.This event occurred 8 times.
 
Additional Manufacturer Narrative
H.6 investigation summary: this complaint is for contaminated phoenix ap id broth (448012) batch 2060448.Customer started seeing purity plates with pseudomonas aeruginosa.Customer did (ap) decontamination and cultured the bag (ref pr #(b)(4)).The bag grew p.Aeruginosa.Customer did additional testing and found the lot to be contaminated.The customer did not provide photos for the investigation.The customer returned one opened and four opened bags of ap id broth batch #2060448.The samples were manually visually inspected and no signs of contamination (cloudy, turbid, etc.) were observed.As part of the investigation, samples from two retention phoenix ap id broth bags of the complaint batch were streaked onto tsa blood agar plates and columbia/can with 5% blood agar plates and placed into the incubator overnight to observe for growth on the plates.No growth was noted on either of the tsa plates or columbia plates for either sample of the phoenix ap id broth.Based on the investigation performed, this complaint is not confirmed.A review of quality notifications revealed no quality notifications generated on the complaint batch.A review of complaints revealed no additional complaints on the complaint batch.Complaint trending was performed, and no trends were identified associated with this defect.Bd id/ast plant quality will continue to monitor for trends and take action as necessary.
 
Event or Problem Description
It was reported while using bd phoenix¿ ap id broth biological contamination was observed.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer that they started seeing purity plates with pseudomonas aeruginosa.Customer did decontamination and cultured the bag.The bag grew p.Aeruginosa.Customer did additional testing and found the lot to be contaminated.This event occurred 8 times.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ AP ID BROTH
Common Device Name
CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15180012
Report Number1119779-2022-01081
Device Sequence Number6222546
Product Code JSG
UDI-Device Identifier30382904480128
UDI-Public0130382904480128
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2023
Device Catalogue Number448012
Device Lot Number2060448
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date08/05/2022
Supplement Report FDA Received Date03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-