BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941816500
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Detachment of Device or Device Component (2907); Activation Failure (3270)
Health Effect - Clinical Codes Stroke/CVA (1770); Foreign Body In Patient (2687)
Date of Event 07/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that distal catheter shaft separation occurred requiring additional intervention.The 90% stenosed target lesion was located in the non-tortuous and calcified proximal right coronary artery.A 5.00 x 16mm synergy xd drug-eluting stent was implanted.However, when the physician removed the stent delivery system (sds), resistance was felt.The physician pulled the device and noticed on fluoroscopy that the sds balloon remained in the artery.The physician attempted to remove the detached sds balloon through snare, re-wire and performed antegrade dissection reentry technique but was unsuccessful.A surgery was scheduled in 30 days to remove the detached component.No patient complications were reported and the patient status was stable.
 
Event or Problem Description
It was reported that distal catheter shaft separation occurred requiring additional intervention.The 90% stenosed target lesion was located in the non-tortuous and calcified proximal right coronary artery.A 5.00 x 16mm synergy xd drug-eluting stent was implanted.However, when the physician removed the stent delivery system (sds), resistance was felt.The physician pulled the device and noticed on fluoroscopy that the sds balloon remained in the artery.The physician attempted to remove the detached sds balloon through snare, re-wire and performed antegrade dissection reentry technique but was unsuccessful.A surgery was scheduled in 30 days to remove the detached component.No patient complications were reported and the patient status was stable.It was further reported that the proximal edge of the stent appeared to be not completely expanded after removal of the sds.
 
Additional Manufacturer Narrative
Updated b5: describe event or problem.Updated h6: device codes.
 
Additional Manufacturer Narrative
Updated b5: describe event or problem.
 
Event or Problem Description
It was reported that distal catheter shaft separation occurred requiring additional intervention.The 90% stenosed target lesion was located in the non-tortuous and calcified proximal right coronary artery.A 5.00 x 16mm synergy xd drug-eluting stent was implanted.However, when the physician removed the stent delivery system (sds), resistance was felt.The physician pulled the device and noticed on fluoroscopy that the sds balloon remained in the artery.The physician attempted to remove the detached sds balloon through snare, re-wire and performed antegrade dissection reentry technique but was unsuccessful.A surgery was scheduled in 30 days to remove the detached component.No patient complications were reported and the patient status was stable.It was further reported that the proximal edge of the stent appeared to be not completely expanded after removal of the sds.It was further reported that the balloon catheter was inflated twice.It was confirmed in the fluoroscopy that all contrast had left the balloon prior to the withdrawal attempt.During the withdrawal of the sds the vacuum was maintained and it was noted that resistance was felt when removing the deployed stent.
 
Event or Problem Description
It was reported that distal catheter shaft separation occurred requiring additional intervention.The 90% stenosed target lesion was located in the non-tortuous and calcified proximal right coronary artery.A 5.00 x 16mm synergy xd drug-eluting stent was implanted.However, when the physician removed the stent delivery system (sds), resistance was felt.The physician pulled the device and noticed on fluoroscopy that the sds balloon remained in the artery.The physician attempted to remove the detached sds balloon through snare, re-wire and performed antegrade dissection reentry technique but was unsuccessful.A surgery was scheduled in 30 days to remove the detached component.No patient complications were reported and the patient status was stable.It was further reported that the proximal edge of the stent appeared to be not completely expanded after removal of the sds.It was further reported that on (b)(6) 2023, the patient underwent angioplasty of the right coronary artery with balloon extraction equipment of the existing balloon with several different wires, without success.Following balloon angioplasty, the patient suffered mini strokes, but was unable to undergo additional imaging due to the balloon catheter being left inside.Following the procedure, the patient expired on (b)(6) 2023.At the time of death, the retained balloon fragments remained in the body.The cause of death revealed acute hypoxic respiratory failure, septic shock and an unknown etiology.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key15180693
Report Number2134265-2022-08123
Device Sequence Number5120860
Product Code NIQ
UDI-Device Identifier08714729981251
UDI-Public08714729981251
Combination Product (Y/N)Y
Initial Reporter StateNV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2023
Device Model NumberH7493941816500
Device Lot Number0026604055
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/17/2022
Supplement Date Received by Manufacturer08/16/2022
09/16/2022
09/08/2023
Initial Report FDA Received Date08/07/2022
Supplement Report FDA Received Date09/06/2022
10/03/2022
09/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER: 6FR; GUIDE CATHETER: 6FR; GUIDE CATHETER: 6FR
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexFemale
-
-