| Model Number |
LSMU1350858 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.(expiry date: 03/2025).
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the stent allegedly broke into two segments.It was further reported that the procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a complaint history review was performed.This is the first complaint reported for this lot number and the lot met all release criteria.Therefore, a device history record review is not required.Investigation summary: the sample was not returned for evaluation.However, a fluoroscopy image was provided for review.The analysis of the image was unable to establish if the image provided represented a fractured stent or if it was of two stents that did not overlap.The result of the investigation is inconclusive for the reported stent fracture issue.The root cause for the reported stent fracture issue could not be determined based upon the available information received from the field communications.Labeling review: the instructions for use for the lifestream product was reviewed and contains the following information relevant to the reported event: indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Warnings: ¿ the covered stent cannot be repositioned after it is deployed.Precautions: ¿ the device should only be used by physicians who are trained in endovascular procedures and are familiar with the complications, side effects and hazards of peripheral vascular interventions.¿ anatomical variances may complicate the procedure; use caution when advancing the endovascular system through tortuous or difficult anatomy.¿ this device has not been tested for use in overlapped conditions with stents or covered stents from other manufacturers.Directions for use: site access and preparation: 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length covered stent size selection.3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation: 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.Introduction of the endovascular system and placement of the covered stent.13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.15.Slowly inflate the endovascular system balloon to nominal pressure, expanding the covered stent.Confirm complete expansion via fluoroscopic visualization.A 15 ¿ 30 second inflation time is recommended.16.After covered stent deployment, apply negative pressure to the balloon until it is fully deflated.Withdraw the delivery system while maintaining negative pressure with the guidewire remaining across the lesion.17.Confirm optimal covered stent wall apposition using standard angiographic techniques.If the covered stent has not achieved full wall apposition, a post dilation with an appropriately sized balloon should be performed.5-8 mm devices may be post-dilated with balloons up to 10 mm in diameter.9-12 mm devices may be post-dilated with balloons up to 12 mm in diameter.D4 (expiry date: 03/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the stent allegedly broke into two segments.It was further reported that the procedure was completed using another device.There was no reported patient injury.
|
| |
|