ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
TRF-605 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: underweight, opening, crease flat and wear abrasion.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is a opening sharp in the posterior side assessed as unidentified (tear) opening.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side exchange from ¿textured to smooth implants due to the patient's concern with the product, and rupture." healthcare professional also reported "brain cancer¿.This event is not device related.Device has been explanted.
|
| |
|