Model Number 1217-01-056
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that intraoperatively, after impacting the cup and screwing in the hole eliminator, the surgeon wanted to correct the position of the cup.In order to insert the cup impactor, the apex hole eliminator should have been removed, which unfortunately was not possible.The surgeon decided to insert the inlay and reposition the hip.As a result of the trial reposition the hip dislocated and the cup was completely changed.No patient harm reported.There was a 30 minute surgical delay.
 
Additional Manufacturer Narrative
(b)(4) depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was received for examination.Visual inspection of the device found it is off axis threaded/mispositioned to the mating device.The reported condition was confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: product description: pinnacle 100 acet cup 56mm.Product code: 121701056, lot number: 3752013.Please provide: 1) quantity manufactured: (b)(4).2) date of manufacture: 05-apr-2022.3) any anomalies or deviations identified in dhr: there were no non-conformances associated with this lot.4) expiry date: 31-mar-2032.5) ifu reference: ifu-0902-00-701.Device history review: a manufacturing record evaluation was performed for the finished device 3752013 121701056, and no non-conformances or manufacturing irregularities related to the malfunction were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 and d10.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 56MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15182821
Report Number1818910-2022-15305
Device Sequence Number17727669
Product Code LPH
UDI-Device Identifier10603295008576
UDI-Public10603295008576
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-056
Device Catalogue Number121701056
Device Lot Number3752013
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2022
Supplement Date Received by Manufacturer08/19/2022
10/25/2022
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date08/22/2022
10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP.; PINN MAR +4 NEUT 32IDX56OD.; PINNACLE 100 ACET CUP 56MM.
Patient SexUnknown
-
-