• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON 64L BD¿ EXTRA LARGE SHARPS COLLECTOR, HINGE TOP; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON 64L BD¿ EXTRA LARGE SHARPS COLLECTOR, HINGE TOP; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305609
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the 64l bd¿ extra large sharps collector, hinge top lid was missing.The following information was provided by the initial reporter: the customer reported that the inner lid was missing.
 
Additional Manufacturer Narrative
The manufacturing location for this product is (b)(4).This site is an oem manufacturing site.(b)(4).Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the 64l bd¿ extra large sharps collector, hinge top lid was missing.The following information was provided by the initial reporter: the customer reported that the inner lid was missing.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval: yes.D.9.Returned to manufacturer on: 14-jul-2022.H.6.Investigation summary: according to the dhr review process; the result showed there were no issues reported like missing slider during the manufacturing process of the part number reported under this customer complaint.A review of the ncmr¿s was performed; the result showed there were no issue reported for missing slider for the same part number throughout the last twelve months.Based on this investigation it was determined that this failure mode is related to the manufacturing process since the lack of slider would be an omission error within the process.The current manufacturing process was reviewed and it was found as capable of detecting missing components, however, omission error could happen during visual inspection.Additionally, this is an isolated issue since there was only two additional complaints received for the last twelve months for the same part number and issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
64L BD¿ EXTRA LARGE SHARPS COLLECTOR, HINGE TOP
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15182950
Report Number2243072-2022-01231
Device Sequence Number17594667
Product Code FMI
UDI-Device Identifier00382903056095
UDI-Public00382903056095
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943575
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number305609
Device Catalogue Number305609
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-