| Model Number |
DRILL SLEEVE, AVN |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022, it was reported by a sales representative via sems that an ar-3515h drill sleeve had an issue, the tip of the sheath bent whenever it touched the bone.This was discovered during use with no impact to the patient.During returned device evaluation, a reportable malfunction was discovered.
|
| |
|
Additional Manufacturer Narrative
|
|
During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed.One unpacked ar-3515h was received for investigation.Visual inspection identified that the distal end of the c8967-01 tube was both dented and broken.The width of the tube was assessed using digital micrometer id: 224 and was identified to meet design specifications.Due to extent of the damage present, the id and od of the breakage site could not be assessed.The cause remains undetermined.
|
| |
|