| Model Number |
TSM-445 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: visual analysis of the returned device identified the device was broken shell and red particles material was found on the shell.A weight test of the device was verified and the device underweight.A microscopic analysis was performed which identified one sharp edge broken in the shell with wear abrasion.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is a sharp edge broken in the shell with wear abrasion in the side posterior assessed as surgical impact opening.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv and rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange from textured to smooth breast implants due to the patient¿s concern with the product." healthcare professional additionally reported " rupture left textured silicone implant" and capsular contracture baker grade iv.Device has been explanted and replaced.
|
| |
|