| Model Number |
DS660HBT |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Sore Throat (2396); Unspecified Respiratory Problem (4464)
|
| Date of Event |
08/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The patient alleged nasal/throat irritation or soreness.There was no report of serious or permanent patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging an issue related to a bipap device's sound abatement foam.The pateint alleged nasal/throat irritation or soreness.There was no report of serious or permanant patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Section h6 updated in this report.Section d7a has been corrected / updated in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received on jul 19 , 2025 and section h10 should be reported as: the manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The patient alleged nasal/throat irritation or soreness.There was no report of serious or permanent patient harm or injury.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.During the evaluation, secondary findings was not observed.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service center.There was 7 error codes found.The third-party service center concludes that they couldn't confirm the customer's allegation and there was no visible foam degradation and unit was scraped.Section h6 updated in this report.
|
| |
|
Search Alerts/Recalls
|