BARD PERIPHERAL VASCULAR, INC. CONQUEST; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ5074
Medical Device Problem Codes Leak/Splash (1354); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.Two photos were provided and reviewed.The first photo shows the introducer sheath inserted in the patient's body.The conquest balloon seen held close to the sheath.The balloon looks partially deflated and bloody.The second photo shows the introducer sheath inserted in the patient's body.The conquest balloon is seen loaded on a guidewire.The guidewire is inserted in sheath while the balloon is held outside the body.The balloon looks partially deflated and bloody.No specific anomalies noted.One video was provided and reviewed.The video shows the introducer sheath inserted in the patient's body.The conquest balloon is seen loaded on a guidewire.The guidewire is inserted in sheath while the balloon is held outside the body.Liquid is seen streaming from the balloon from an unclear source as the balloon is being inflated.No other anomalies noted.Therefore, the investigation is inconclusive for the reported balloon rupture as no evidence of rupture can be determined.The investigation is confirmed for the identified leak as water is seen streaming from the balloon from the provided video.However, the specific cause of the leak cannot be determined.A definitive root cause for the reported balloon rupture and identified leak could not be determined based upon the provided information.Labelling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 05/2024).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15184078
Report Number2020394-2022-00647
Device Sequence Number5848105
Product Code DQY
UDI-Device Identifier00801741062957
UDI-Public(01)00801741062957
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ5074
Device Catalogue NumberCQ5074
Device Lot NumberREFT3285
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2022
Initial Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
-
-