MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 9L 14MM, VE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 9L 14MM, VE Back to Search Results
Model Number 341-14-709
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 20 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported device was the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR 3D KNEE
Common Device Name
EMPOWR 3D KNEETM INS, 9L 14MM, VE
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15184443
Report Number1644408-2022-01025
Device Sequence Number6389855
Product Code JWH
UDI-Device Identifier00888912167062
UDI-Public(01)00888912167062
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/23/2026
Device Model Number341-14-709
Device Catalogue Number341-14-709
Device Lot Number133T1062
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-