MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565030
Medical Device Problem Codes Difficult to Remove (1528); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to boston scientific corporation on july 18, 2022 that a wallflex duodenal stent was to be implanted to treat an approximately 5 cm malignant stricture in the duodenum during an esophagogastroduodenoscopy (egd) with duodenal stent placement procedure performed on (b)(6) 2022.The patient's anatomy was tortuous prior to stent placement.During the procedure, the wallflex duodenal stent was deployed in the correct location.During removal of the wallflex duodenal delivery system, the delivery system got caught on the wallflex duodenal stent.As the physician continued to withdraw the delivery system, the delivery system pulled the wallflex duodenal stent and eventually the wallflex duodenal stent moved out of place.The wallflex duodenal stent was removed from the patient with rat- tooth grasping forceps and a 22 x 90 wallflex duodenal stent was implanted to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the wallflex duodenal stent was disposed and will not be returned for evaluation; therefore, a device analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL
Common Device Name
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15186628
Report Number3005099803-2022-04294
Device Sequence Number5847661
Product Code MUM
UDI-Device Identifier08714729456506
UDI-Public08714729456506
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K062750
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2023
Device Model NumberM00565030
Device Catalogue Number6503
Device Lot Number0027741039
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2022
Initial Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-