| Model Number |
M00565030 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on july 18, 2022 that a wallflex duodenal stent was to be implanted to treat an approximately 5 cm malignant stricture in the duodenum during an esophagogastroduodenoscopy (egd) with duodenal stent placement procedure performed on (b)(6) 2022.The patient's anatomy was tortuous prior to stent placement.During the procedure, the wallflex duodenal stent was deployed in the correct location.During removal of the wallflex duodenal delivery system, the delivery system got caught on the wallflex duodenal stent.As the physician continued to withdraw the delivery system, the delivery system pulled the wallflex duodenal stent and eventually the wallflex duodenal stent moved out of place.The wallflex duodenal stent was removed from the patient with rat- tooth grasping forceps and a 22 x 90 wallflex duodenal stent was implanted to complete the procedure.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The complainant indicated that the wallflex duodenal stent was disposed and will not be returned for evaluation; therefore, a device analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|