| Model Number |
DSL18P |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Our investigation is ongoing.A follow up report will be submitted when the investigation is complete.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Catheter found to allow blood infused in one lumen to flow through the other lumen.This resulted in recirculation of dialysate and inadequate dialysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in the incident was not returned for evaluation.Additional information and clarification was requested but not provided.The contract manufacturer conducted a review of the manufacture records for the lot number provided.A record review of the device lot numbers showed no nonconformities or other abnormalities.There were no deviations or process changes in the manufacturing and/or materials with the part numbers.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have revealed this type of failure if it had existed at the time of manufacture.Without an evaluation of the device a root cause cannot be determined.
|
| |
|