| Catalog Number |
0684-00-0293 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complete reporter name: dr.(b)(6).Complete event site name - (b)(6).Complete event site address - (b)(6).Event site postal code - (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that after approximately 30 minutes of intra-aortic balloon (iab) therapy, the console generated a leak in iab circuit alarm and blood was seen in the helium tubing.After removal, it was found that the inner lumen was broken.A new iab was inserted to continue therapy.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint (b)(4).
|
| |
|