• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 40375
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that tip detachment occurred.Vascular access was obtained via up-and-over technique.The 100% stenosed target lesion was located in the non-tortuous and moderately calcified superficial femoral artery.A 014/300/sst platinum plus guidewire and a non-boston scientific (bsc) support catheter were advanced into the lesion.However, during the procedure, the guidewire tip fractured.The detached fragment was retrieved by snaring and arterial exploration.The procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: unit returned in a generic plastic bag.The unit returned has tip damaged, as part of overall visual revision.The wire returned was inspected and it is possible to observe the spring tip is unraveled and detached and corewire bent and detached.The outer dimensions were found within specification.The device was sent for material analysis, and evidence of necking, cone fracture surface and irregular appearance on the top of the coil suggest that it experienced a tensile load.The curved mechanical marks also suggest that it experienced twisting or bending actions.
 
Event or Problem Description
It was reported that tip detachment occurred.Vascular access was obtained via up-and-over technique.The 100% stenosed target lesion was located in the non-tortuous and moderately calcified superficial femoral artery.A 014/300/sst platinum plus guidewire and a non-boston scientific (bsc) support catheter were advanced into the lesion.However, during the procedure, the guidewire tip fractured.The detached fragment was retrieved by snaring and arterial exploration.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM PLUS
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15188065
Report Number2124215-2022-26174
Device Sequence Number17618640
Product Code DQX
UDI-Device Identifier08714729486886
UDI-Public08714729486886
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K945379
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2025
Device Model Number40375
Device Catalogue Number40375
Device Lot Number0029321270
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer09/04/2022
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SUPPORT CATHETER: TRAILBLAZER; SUPPORT CATHETER: TRAILBLAZER
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient RaceBlack Or African American
-
-