| Model Number |
358764 |
| Medical Device Problem Code |
Ambient Noise Problem (2877)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Event or Problem Description
|
|
Boston scientific provided the following information: non-physiologic tachycardia events with noise (no syncope) were observed on both channels.Patient is device dependent.Maximum length of event was 4 seconds.Lead remains implanted at this time.Should additional information be received, this file will be updated.
|
| |
|