BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA PRO 1.5/15; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER Back to Search Results
Model Number 393299
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
A pantera pro coronary dilatation catheter was chosen for treatment of a moderately calcified lesion (95 percent stenosis degree) in the moderately tortuous proximal rca.When removing the transportation wire, the balloon was stuck to the wire and tore off the delivery system.It was not recognized, and the device was introduced into the guiding catheter.Because the balloon was not seen in the imaging the device was removed.Another balloon was used to complete the intervention.Udi (b)(4).
 
Additional Manufacturer Narrative
The returned product was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The instrument was returned fractured in two parts.The distal fragment (including balloon) is stuck on the transportation wire and partially covered by the balloon protector.The balloon protector has been pulled over the loop at distal end of transportation wire.The fracture site is located at the proximal balloon weld.The inner shaft is severely deformed.Also, the device tip is severely deformed and has necked near the tip weld, both indicating application of a high tensile force.Backscatter electron imaging and energy-dispersive x-ray spectroscopy of both the device tip inner surface and transportation wire outer surface revealed contrast medium residue which may have contributed to the complaint event.Moreover, simulations have shown that a similar damage pattern as observed in the returned instrument could be reproduced by wrong handling (e.G.Grabbing the balloon protector in the tip area when pulling).Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.The transportation wire is moved several times during production.As part of the inner packaging process each instrument is inspected for damages (100 percent control) and the position of the transportation wire is checked and corrected, if necessary.Based on the conducted investigations, no manufacturing or material related root cause could be determined.Please note that the ifu advises the user to pull the plunger of the inflation/deflation device and aspirate air from the catheter prior to introduction, and to not use the catheter if a vacuum cannot be held, as this indicates a system failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA PRO 1.5/15
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15188171
Report Number1028232-2022-04008
Device Sequence Number6612480
Product Code LOX
UDI-Device Identifier07640130437036
UDI-Public(01)07640130437036(17)2210
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K160985
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2022
Device Model Number393299
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10193497
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-