• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT KNEE COMPONENTS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT KNEE COMPONENTS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number TRULIANT TIB IMP PSC INSERT SZ 2, 10MM
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint effusion (4849)
Date of Event 07/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: truliant ps cem fem ps cem left sz 2 (cat# 02-020-11-0220), truliant tib fit tray cem sz 2f / 2t (cat# 02-022-45-2020), advanced patella 29m 3 peg implant (cat# 200-07-29), holding pin mini sharp point 4 pk (cat# 201-78-15).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by the exactech truliant knee clinical study; approximately two years post initial left tka, the 89 y/o female patient had pain in left knee.The event was resolved as patient had joint effusion and cortisone injection.
 
Additional Manufacturer Narrative
As reported by the exactech truliant knee clinical study; approximately two years post initial left tka for osteoarthritis, the (b)(6) had pain in left knee.The event was resolved as patient had joint effusion and cortisone injection.No additional information.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the subsequent pain cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.These devices are used for treatment not diagnosis.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d1, d4 (remove all from initial report-suspect device unknown), g4(remove 501k from initial report-unknown), h6: clinical code, medical device problem code, component code, type of investigation and investigation conclusion.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT KNEE COMPONENTS
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15188290
Report Number1038671-2022-00900
Device Sequence Number5847187
Product Code JWH
UDI-Device Identifier10885862304902
UDI-Public10885862304902
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171045
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTRULIANT TIB IMP PSC INSERT SZ 2, 10MM
Device Catalogue Number02-022-44-2010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer08/23/2022
11/28/2025
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date09/23/2022
12/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
Patient SexFemale
Patient Weight57 KG
-
-