| Model Number |
FOLATE G3 |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Codes |
Conjunctivitis (1784); Chemical Exposure (2570); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Occupation is roche field application specialist (fas).Roche medical assessment states: tobramycin ophthalmic ointment is a topical antibiotic indicated where superficial bacterial ocular infection or a risk of bacterial ocular infection exists.Eye tissue damage directed from outside by a splash of irritant substance affect primarily conjunctiva, cornea and anterior cavity.The initial injury may promptly get complicated by a secondary infection.Preventive administration of topical antibiotics is a reasonable medical measure widely followed.Benefit of preventive antibiotics outweighs the risk of complications due to infection.Product labeling provides the method of prevention: "wear protective gloves/protective clothing/eye protection/face protection." the investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of an injury to the technician while checking for film on the elecsys folate iii (folate iii) reagent of a cobas e801 module serial number (b)(4).A film alarm occurred and the technician reportedly removed the reagent from the instrument to check for bubbles.While checking the reagent, the technician allegedly splashed the reagent in her eye.The technician was wearing gloves and a lab coat; she was not wearing face or eye protection.The technician flushed her eye with water and went to the urgent care center where they reportedly suggested she see an eye specialist.The technician reportedly saw an ophthalmologist who provided her with tobramycin eye ointment.The ophthalmologist allegedly discovered "damage" to her eye but did "not believe it is permanent." clarification was requested related to the alleged "damage" to the eye but the doctor "did not elaborate" on this statement.The technician is currently reportedly "doing ok" and is back at work.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined the event was due to a product handling issue at the customer site.
|
| |
|