• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS Back to Search Results
Model Number AR40E
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section age or date of birth, weight and ethnicity: unknown/ not provided.If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted.Telephone number, (b)(6).The intraocular lens (iol) was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Event or Problem Description
Customer reported a torn sensar ar40e intraocular lens (iol).No patient injury reported.No further information provided.
 
Additional Manufacturer Narrative
Corrected data: section b3 date of event: additional information was received through follow-up activity.The correct date of event was provided as 06/06/2022.Additionally, the following information was also received: section a2: patient's date of birth: (b)(6) 1952.Section a3: patient's gender: female.Section a4: ethnicity: hispanic.Section e1: doctor's name: (b)(6).Doctor's email: (b)(6).Doctor's phone number:(b)(6).Section e2: health professional: yes.Section e3: occupation: physician.Customer had a problem mounting lens to cartridge.The lens or cartridge had contact with the patient.The affected eye was the left eye.Medical intervention required: doctor attempted scleral fixation of the lens.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9.Device available for evaluation? yes.Returned to manufacturer on oct 7, 2022.Section h3.Device evaluated by manufacturer? yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint lens was received coated in viscoelastic residue and with both haptics detached.The lens was cleaned and, no further issues could be identified.The complaint issue "iol torn¿ was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no other complaints were received for this po.Therefore, no escalations are required.Conclusion: as a result of the investigation, there is no indication of a product malfunction or product quality deficiency.No nonconformity report, documentation or labeling changes, and escalations are required.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15189164
Report Number3012236936-2022-02099
Device Sequence Number5541844
Product Code HQL
UDI-Device Identifier05050474502192
UDI-Public(01)05050474502192(17)220726
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/26/2022
Device Model NumberAR40E
Device Catalogue NumberAR40E00245
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer09/30/2022
10/19/2022
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date10/29/2022
11/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient EthnicityHispanic
-
-