• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 ULTRA 4MM X 8CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 ULTRA 4MM X 8CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035424080
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during an endovascular embolization procedure, physician used the coils but the first coil was unable to detach.The subject coil was used as the second coil and detached successfully but it was floating freely inside the patient's anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No additional information available.
 
Event or Problem Description
It was reported that during an endovascular embolization procedure, physician used the coils but the first coil was unable to detach.The subject coil was used as the second coil and detached successfully but it was floating freely inside the patient's anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No additional information available.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date ¿ added.Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that the coil did detach but was free floating inside patient.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this event of 'main coil migration¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 ULTRA 4MM X 8CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15189253
Report Number3008881809-2022-00391
Device Sequence Number5977242
Product Code HCG
UDI-Device Identifier04546540675927
UDI-Public04546540675927
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2024
Device Model NumberM0035424080
Device Catalogue NumberM0035424080
Device Lot Number22983535
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer09/25/2022
Initial Report FDA Received Date08/08/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INZONE POWER SUPPLY (STRYKER)
Patient SexUnknown
-
-