| Model Number |
RT225 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in china, via a fisher & paykel healthcare (f&p) field representative, that the inspiratory limb of an rt225 infant ventilator circuit was found cracked.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint (b)(4) infant bias flow breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand for investigation, where it was visually inspected.Results: visual inspection revealed a cut in the inspiratory limb of the subject circuit.It was also noted that the (b)(4) packaging bag was found cut open.Conclusion: we were unable to determine what may have caused the reported event.However, the inspiratory limb appeared to be cut with a sharp object.All (b)(4) infant bias flow breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the (b)(4) infant bias flow breathing circuit state the following: "check all connections are tight before use."."perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient."."ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient."."visually inspect breathing sets for damage before use and replace if damaged.".
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in china, via a fisher & paykel healthcare (f&p) field representative, that the inspiratory limb of an rt225 infant bias flow breathing circuit was found cracked.There was no patient involvement.
|
| |
|