The catalog number identified has not been cleared in the us but is similar to the groshong 5.5f single-lumen cv catheter that are cleared in the us.The pro code and 510 k number for the groshong 5.5f single-lumen cv catheter are identified.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the device was not returned for evaluation.The complaint is confirmed for the identified product identification/marketing information issue, as the investigation confirmed an outdated german brochure, a german website, and certain eu jde german-russian descriptive information incorrectly indicated the device contained an introducer.The marketing team found no other cvc products to be affected.However, the problem has only been discovered in eu jde for 7711550ce at this time.However, the investigation is unconfirmed for the reported issue as no evidence related to missing component issue was found.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Expiry date: 07/2022.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device not returned.
|