W.O.M. WORLD OF MEDICINE GMBH AQUILEX FLUID CONTROL SYSTEM AQL; HYSTEROSCOPIC INSUFFLATOR, FLUID MONITORING SYSTEM AND TUBE SETS Back to Search Results
Model Number AQL-100PBS
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
We have been informed of the followind event: "part description: aql pbs serial number or, lot number of reported item: 1911ce1150 , date of procedure: (b)(6) 2022, date product support or hologic employee notified of issue reported: (b)(6) 2022, issue reported: customer called tm explaining that during a procedure, the deficit jumped from 2000mls down to 200mls.Tm stated the fluid deficit test passed after checking post op.The customer stated 19+ saline bags were used in the 2-hour procedure.The customer left all fluid bags on the hooks and did not lift any bags while replacing during procedure.The customer also claimed the deficit went negative at some point during the procedure.Troubleshooting performed: no information provided how did the procedure complete: despite the inaccuracies of the deficit, the procedure completed without injury.Patient status: no injury reported.Replacement requested: n.Letter of investigation requested: n.Is product returning for investigation: n.Site contact information: na.Additional information: apps spec confirmed manual count was 1500mls and pressure was set to 80mmhg throughout the procedure.The mean arterial pressure (map) is unknown.It is uncertain how the deficit would change from large amount to small amount to negative throughout procedure.Nothing was draped on the scales at any time.Case closed.Patient/user impact no impact reported".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUILEX FLUID CONTROL SYSTEM AQL
Common Device Name
HYSTEROSCOPIC INSUFFLATOR, FLUID MONITORING SYSTEM AND TUBE SETS
Manufacturer (Section D)
W.O.M. WORLD OF MEDICINE GMBH
salzufer 8
berlin, 10587
GM  10587
Manufacturer (Section G)
W.O.M. WORLD OF MEDICINE GMBH
alte poststrasse 11
ludwigsstadt, 96337
GM   96337
Manufacturer Contact
ellen giese
salzufer 8
berlin, 10587
GM   10587
MDR Report Key15189747
Report Number3002914049-2022-00006
Device Sequence Number5851894
Product Code HIG
UDI-Device Identifier04056702006914
UDI-Public04056702006914
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K172040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAQL-100PBS
Device Catalogue NumberAQL-100PBS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Initial Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1