| Catalog Number |
UE514010300 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the cart is reading full when its actually empty.The event timing was outside of surgery.There was no harm or delay reported.No adverse events were reported as a result of this malfunction.
|
| |
|
Event or Problem Description
|
|
There is no additional event information available.
|
| |
|
Additional Manufacturer Narrative
|
|
The unit was reading cylinder 1 as full when it was empty.The level sensor was replaced.Dhr was reviewed and no discrepancies related to the reported event were found.The level sensor is required to read and report fluid level in the cylinders in order for the unit to function as intended.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|