• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U35130108
Medical Device Problem Codes Leak/Splash (1354); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse 035 pta dilatation catheter was received for evaluation.During visual evaluation, a tear was noted on the catheter shaft.No kinks noted to the catheter shaft and no anomalies to the luers/y-body.During functional testing, an in-house presto inflation device was used to inflate the balloon and water was noted leaking from the tear on the catheter.One video was provided and reviewed.The video shows the ultraverse 035 catheter.During irrigation of the device, leak is noted at one of the geo markings of the catheter.However the specific cause of the leak could not be identified in the video.No other anomalies noted.Two photos were provided and reviewed.The first photo shows the front of the outer packaging of the ultraverse 035 catheter with the labeling corresponding to the device.The second photo shows the back of the outer packaging.Two ultrasound images were provided and reviewed.A vessel is shown in each image.It appears that either these are two different vessels or visualized at different planes, however there is no context to the images provided nor translation of the report below the images.The details and settings of the ultrasound cannot be determined from the provided images due to poor image quality.Therefore, the investigation is confirmed for the reported leak as liquid was noted leaking from the catheter shaft during irrigation of the device, from the provided video and during functional testing of the returned sample.The investigation is confirmed for the identified catheter tear as a tear was noted on the catheter shaft during visual evaluation of returned device.A definitive root cause for the alleged leak and identified catheter tear could not be determined based upon the available information.Labelling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 08/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the device allegedly leaked from the marking band of the balloon body.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15190898
Report Number2020394-2022-00651
Device Sequence Number12265339
Product Code LIT
UDI-Device Identifier00801741094729
UDI-Public(01)00801741094729
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU35130108
Device Catalogue NumberU35130108
Device Lot NumberCMFW0203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2022
Initial Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight70 KG
-
-