MEDOS INTERNATIONAL SÃ RL CH COUGAR LS LATERAL CAGE SYSTEM LORDOTIC CAGE 18 X 12 X 55MM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Model Number |
189155512 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, the cougar ls lateral lordotic cage broke during impaction when implanted.Looking at x-ray, it was found one of the marking pins were angled, and the implant did not span the disc space.Surgeon removed the broken implant, cleaned out the disc space and inserted a new implant.New implant went right in.The procedure was completed successfully with no reported adverse patient impact.This report is for a cougar ls lateral cage system lordotic cage 18 x 12 x 55mm.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Part # 516769-70 lot # 189155512 supplier: (b)(4) ¿ batch1: lot qty of (b)(4) units were released on 03 march 2021 with no discrepancies.¿ batch2: lot qty of (b)(4) units were released on 16 feb 2021 with no discrepancies.¿ batch3: lot qty of (b)(4) units were released on 21 jan 2021 with no discrepancies.¿ batch4: lot qty of (b)(4) units were released on 08 april 2020 with no discrepancies.¿ batch5: lot qty of (b)(4) units were released on 06 april 2020 with no discrepancies.¿ batch6: lot qty of (b)(4) units were released on 18 dec 2019with no discrepancies.¿ batch7: lot qty of (b)(4) units were released on 04 dec 2019with no discrepancies.¿ batch8: lot qty of (b)(4) units were released on 21 nov 2019 with no discrepancies.¿ batch9: lot qty of (b)(4) units were released on 28 oct 2019 with no discrepancies.¿ batch10: lot qty of (b)(4) units were released on 28 oct 2019 with no discrepancies.¿ batch11: lot qty of (b)(4) units were released on 27 sep 2019 with no discrepancies.¿ batch12: lot qty of (b)(4) units were released on 10 sep 2019 with no discrepancies.¿ batch13: lot qty of (b)(4) units were released on 05 sep 2019 with no discrepancies.¿ batch14: lot qty of (b)(4) units were released on 26 aug 2019 with no discrepancies.¿ batch15: lot qty of (b)(4) units were released on 14 aug 2019 with no discrepancies no ncrs were generated during production.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that cougar ls 18x12x55 lordotic was observed to have broken in two section, the fragment was received in the evidence provided.A dimensional inspection for the cougar ls 18x12x55 lordotic was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the cougar ls 18x12x55 lordotic would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|