This report is being submitted to report additional information.The product was returned, and the reported event could not be verified as the exact details of event were unverifiable.Based on the evaluation, devices malfunction did not occur.Additionally, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported devices that did or could cause or contribute to the reported event.Monthly post market trending review identified no actionable trends or corrective actions for the reported event or devices.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.Therefore, based on the available information, the product was within specifications and conforming when it left zimvie.Dhr review was completed for the subject lot numbers.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot numbers were inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot numbers for similar events and no other complaint was identified.Functional testing to recreate the reported event; verified the devices engaged, tightened, retained and disengaged as normal.No malfunction identified.A definitive root cause could not be determined.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
|