Model Number CAX600H12W
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Headache (1880)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The patient alleged bad headaches while using the device.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The dreamstation bipap pro device was returned to a third-party service center for further evaluation.During evaluation, sound abatement foam degradation was not observed.The service center retrieved and reviewed the device's error logs.There were no errors found.The third-party service center could not confirm the customer's allegation and there was no visible foam degradation.The unit was scrapped after device evaluation was completed.Sections g1, h2, h3, and h6 have been updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP PRO
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7243512041
MDR Report Key15191201
Report Number2518422-2022-69177
Device Sequence Number5602834
Product Code BZD
UDI-Device Identifier00606959034716
UDI-Public00606959034716
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 10/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCAX600H12W
Device Catalogue NumberCAX600H12W
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer02/27/2025
Initial Report FDA Received Date08/09/2022
Supplement Report FDA Received Date10/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-